--- # Protocol metadata id: PROT-0023 name: Counting Cells Manually Protocol version: 1.0 description: Protocol for manual cell counting using a hemocytometer and trypan blue author: Lab Staff created: 2025-05-06 last_updated: 2025-05-07 category: cell-culture # Materials required materials: - name: Cell suspension preparation: Freshly harvested with trypsin/dissociation reagent temperature: Room temperature - name: Trypan Blue solution concentration: 0.4% in PBS storage: Room temperature - name: Microcentrifuge tubes type: 1.5 mL sterile - name: Hemocytometer type: Standard with coverslip - name: PBS concentration: 1X, sterile use: Optional for dilution if cell concentration is too high # Equipment required equipment: - name: Microscope type: Inverted or upright with 10x objective - name: Pipettes range: 2-200 μL - name: Calculator use: For cell counting calculations - name: Counter (optional) use: Manual cell counter or clicker - name: Tissue wipes use: For cleaning hemocytometer # Protocol steps steps: - step: 1 action: "Prepare cell suspension" details: "After harvesting cells with trypsin or other dissociation reagent, mix the cell suspension well by gentle pipetting" - step: 2 action: "Sample the cell suspension" details: "Transfer 10–20 μL of the cell suspension into a microcentrifuge tube" - step: 3 action: "Add trypan blue" details: "Add an equal volume of 0.4% Trypan Blue solution to the cell suspension and mix gently" - step: 4 action: "Load hemocytometer" details: "Apply 10 μL of the cell/trypan blue mixture to the edge of the coverslip on the hemocytometer chamber and allow to fill by capillary action" - step: 5 action: "Wait for cells to settle" details: "Allow cells to settle for 10-30 seconds before counting" - step: 6 action: "Count cells" details: "Count live cells (unstained, bright) and dead cells (blue-stained) in the four outer corner squares of the hemocytometer grid" - step: 7 action: "Calculate cell concentration" details: "Apply the formula: Cells/mL = Average count per square × Dilution factor × 10⁴" - step: 8 action: "Calculate viability percentage" details: "Viability (%) = [Live cell count ÷ (Live cell count + Dead cell count)] × 100" - step: 9 action: "Calculate total viable cells" details: "Total viable cells = Cell concentration (cells/mL) × Total volume of cell suspension (mL) × Viability (%)/100" # Calculation formulas calculations: - calculation: "Cell concentration" formula: "Cells/mL = Average count per corner square × Dilution factor × 10⁴" example: "25 cells/square average × 2 (dilution) × 10⁴ = 5 × 10⁵ cells/mL" - calculation: "Cell viability" formula: "Viability (%) = [Live cell count ÷ (Live cell count + Dead cell count)] × 100" example: "80 live cells ÷ (80 live + 20 dead) × 100 = 80% viability" - calculation: "Total viable cells" formula: "Total viable cells = Cells/mL × Total volume (mL) × (Viability % ÷ 100)" example: "5 × 10⁵ cells/mL × 10 mL × 0.8 = 4 × 10⁶ total viable cells" # Critical parameters critical_parameters: - parameter: "Cell mixing" details: "Ensure thorough but gentle mixing to get a uniform suspension without damaging cells" - parameter: "Counting area" details: "Count cells touching the top and left lines of each corner square, but not the bottom or right lines" - parameter: "Counting time" details: "Count within 3-5 minutes of trypan blue addition; longer exposure can lead to false positives" # Troubleshooting troubleshooting: - problem: "Too many cells to count" solution: "Dilute sample further with PBS and repeat, adjusting dilution factor in calculations" - problem: "Too few cells to count" solution: "Concentrate sample by centrifugation and resuspend in smaller volume" - problem: "Air bubbles in chamber" solution: "Clean and dry hemocytometer and coverslip, then reload carefully" - problem: "Uneven cell distribution" solution: "Mix cell suspension more thoroughly before sampling" # Safety considerations safety: ppe: "Lab coat and gloves required" hazards: "Trypan blue is potentially carcinogenic; handle with care and dispose of properly" # Quality control quality_control: - check: "Cell number" criteria: "Count at least 100 cells total for statistical reliability" - check: "Chamber loading" criteria: "Ensure chamber is not under- or overloaded; cells should be in a single plane" - check: "Replicate counts" criteria: "Count both chambers of hemocytometer; values should be within 10% of each other" # References references: - "Strober W. (2015) Trypan Blue Exclusion Test of Cell Viability. Curr Protoc Immunol. 111:A3.B.1-A3.B.3" - "Louis KS, Siegel AC. (2011) Cell viability analysis using trypan blue: manual and automated methods. Methods Mol Biol. 740:7-12" # Notes notes: | - Trypan blue stains dead cells blue, while live cells remain unstained - For accurate results, count at least 100 cells total - Dilution factor must be accounted for in final calculations - If cell clumps are present, they may indicate incomplete dissociation - Hemocytometer and coverslip must be clean and free of scratches - The hemocytometer chamber depth is 0.1 mm and each corner square has an area of 1 mm² - The multiplier 10⁴ converts the count to cells per mL (1 cm³ = 1000 mm³ = 1 mL) - For primary cells or delicate cell lines, consider automated counting methods ---